Installation Procedures
4-2 BioRobot EZ1 DSP Workstation User Manual 02/2007
workstation. To preserve this protection feature, do not
operate the workstation from an AC power outlet that has no
ground (earth) connection.
Installation of AC power cord
Connect one end of the AC power cord to the socket located
on the rear of the BioRobot EZ1 DSP, and the other end to
the AC power outlet.
The fuse box of the BioRobot EZ1 DSP is located under the
socket for the power cord and contains 2 fuses which are
labeled as follows:
110–120 V — a 6.3 A (250 V) fuse compatible with
110–120 V power supplies
220–240 V — a 3.15 A (250 V) fuse compatible with
220–240 V power supplies
Before connecting the instrument to the power supply, you
may need to rotate the fuse box in order to select the correct
fuse. A fuse is correctly selected when its label is readable
and closer to the bottom of the instrument.
Selecting the Correct Fuse for a
110–120 V Power Supply