Preventive Maintenance
BioRobot EZ1 DSP Workstation User Manual 02/2007 6-5
6. Wipe the O-rings of the tip adapters with tissue.
6.3 Weekly maintenance procedure
Perform the daily maintenance procedure before you
perform the weekly maintenance procedure.
To maintain good contact between tip adapters and filter-tips
and to prevent liquid leaking from the tips, grease the
O-rings of the tip adapters every week:
1. Apply a small amount of silicon grease to the end of a
filter-tip.
2. Apply the silicon grease to the surface of the O-rings.
3. Place the tip onto the pipettor head, and rotate the tip on
the pipettor head to distribute the silicon grease evenly.
Note: The filter-tips should sit flush against the upper
white plastic bar if the O-rings are properly greased.
There should not be a gap. Excess or insufficient grease
can affect the performance of the BioRobot EZ1 DSP.
O-ring
Gap
Seitenansicht 54