Sharp CD-M4000 Betriebsanweisung Seite 114

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 204
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 113
System Status
4-24 72A-1230 Rev. F
July 22, 2005
Figure 4.3 System Status Test Results
This test is passed when every item’s status is reported as Enabled. Any other
result has the following implications:
Guard Status
Disabled
Critical: This condition must be corrected before testing
continues.
Corrective Procedure: Replace the Guard Board and run the Field
Recalibration Test.
The consequences of continued testing in the uncorrected state are:
Calibration Verification cannot be performed
Measurement error uncertainty (2% Capacitance relative and 0.1%
Power Factor absolute)
Guard failure in Subsystem Diagnostics
Main Reference
Status Disabled
Critical: This condition must be corrected before testing
continues.
Corrective Procedure: Replace the Main Reference Assembly and run the
Field Recalibration Test.
The consequences of continued testing in the uncorrected state are:
Calibration Verification cannot be performed
Seitenansicht 113
1 2 ... 109 110 111 112 113 114 115 116 117 118 119 ... 203 204

Kommentare zu diesen Handbüchern

Keine Kommentare