A-104 Color Measurement Instruments
Sharp AR-C360P EFI Color Guide - 104
9. If you have not already done so, remove the shipping spacer from
the measurement page slot.
10. Insert the end of the Color Reflection Reference strip into the
calibration strip entrance on the DTP41. Center the arrow on the
strip below the alignment mark (see the diagram on page A-98),
and slide the strip into the DTP41 until you feel the resistance of
the feed mechanism.
11. Click Read Reference in the Instrument Calibration dialog box.
The DTP41 pulls the calibration strip through automatically.
A message appears in the Status area when the calibration is
finished.
12. Click Done.
13. Close the Utilities window.
This completes DTP41 calibration.
When the DTP41 requires calibration, your computer prompts you.
When the prompt message appears, click Calibrate Now and
follow the instructions, starting with step 8 on page A-103.
After calibrating the DTP41, calibrate the Fiery (see page 2-57).