STP-200/300/400 Series Instruction Manual
4.4.2 How to Connect the Cables
When connecting/disconnecting cables, always power OFF the STP
pump (Switch the breaker "OFF").
Failure to do so may result in electric shock or product damage.
With each connector, align the position of the guide key and insert
the pin vertically so as not to bend it.
If the pin is bent, not only may the connector not function normally,
but may make the pins contact each other, resulting in a malfunction.
Lock and securely tighten each connector and screw.
DO NOT apply voltage to each pin and DO NOT cause any
short-circuiting between pins.
Connect each cable securely. DO NOT place heavy objects on the
cables nor bend them excessively.
Support each cable so as not to apply direct force to the connectors
or terminals.
If any problem occurs in cables, connectors or terminals, the STP
pump may not function normally.
1) Connecting the Ground Cable
Connect the ground cable (yellow/green) between the ground terminal of the STP
pump and the ground terminal of the STP control unit.
When the resistance between the ground terminals is lower than 0.1 , it is not
necessary to connect the ground cable after installing the STP pump and the STP
control unit.
First, connect the ground cable. Next, connect other cables.
When the resistance between the ground terminals is over 0.1 ,
always connect the ground cable. Failure to do so may result in
electric shock.
4-11