STP-200/300/400 Series Instruction Manual
6.1.3 Excessive Vibration
When serious vibration or mechanical shock causes the rotor to touch the touch
down bearing (due to external vibration/impact, intrusion of atmosphere or
foreign materials into the STP pump or rotor imbalance ) the STP pump enters
the STOP operation.
The "EMERGENCY OPERATION" lamp lights.
6.1.4 Inverter Overload
When the STP pump does not attain the rated speed within approx. 30 minutes
after starting or when the ACCELERATION state remains unchanged during
operation for approx. 30 minutes, the inverter stops and the STP pump stops as
well. The "FAILURE" lamp lights.
6.1.5 Overheating Inside the STP Pump
When the temperature of the motor inside the STP pump exceeds 90°C due to an
abnormal baking temperature or overload operation, the STP pump decelerates
down to 40,000 rpm.
The "OVER TEMPERATURE" lamp lights.
Further, when the temperature inside the STP pump exceeds 100°C, the STP
pump enters the STOP operation.
The "OVER TEMPERATURE" and "FAILURE"lamps light.
6.1.6 Overheating Inside the STP Control Unit
When the temperature inside the STP control unit (at the heat sink) exceeds
75 ℃ due to a failure in the air cooling fan, external heat source, and others, the
STP pump enters the STOP operation.
The "FAILURE" lamp lights.
6.1.7 Overspeed
When the rotational speed of the STP pump exceeds 52,500 rpm due to a failure
in the inverter, the STP pump enters the STOP operation.
The "FAILURE" lamp lights.