Sharp SYSTEM 8/10 Plus Bedienerhandbuch Seite 182

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 403
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 181
Calibration
Calibration Procedure
4-14 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
In the Main Menu, choose Service > Neph Calibration > Neph Temp.
CALIBRATE NEPH TEMP:
NEPH TEMP: 0.0
o
C
OFFSET: 0.0
o
C
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Nephelometer Source Level screen allows the user to view and
calibrate the source level. This should be within 60-70 mA. If it is outside
of this range, please adjust the % SRC Level until the IRED reads 65 mA
In the Main Menu, choose Service > Neph Calibration > Neph Src
Level.
SOURCE LEVEL SET:
IRED: 65 mA
SRC LEVEL: 32 %
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Nephelometer Source
Level
Seitenansicht 181
1 2 ... 177 178 179 180 181 182 183 184 185 186 187 ... 402 403

Kommentare zu diesen Handbüchern

Keine Kommentare