C-Link Protocol Commands
Diagnostics
B-24 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
baro pres
This command reports the current barometric pressure, in mmHg. The
following example reports that the barometric pressure is 722 mmHg.
Send: baro pres
Receive: baro pres 722.0 mmHg
board temp
This command reports the current board temperature, in degrees C. The
following example reports that the board temperature is 15 °C.
Send: board temp
Receive: board temp 15.0 degC
crn
This command reports the current radon concentration for the detector.
The following example reports that the radon concentration is 1.2 Bq/m
3
.
Send: crn
Receive: crn 1.20 Bq/m3
det status
This command reports the current mass for the detector. The sequence is:
α, β, β
C
, β
REF
. Each value is separated by a space.
Send: det status
Receive: det status 1 9657 9600 4623
diag volt det
This command reports the diagnostic voltage measurements on the
detector board. The voltage is positive 5.
Send: diag volt det
Receive: diag volt det 4.9
diag volt iob
This command reports the diagnostic voltage measurements on the I/O
expansion board. The sequence of voltages is: Positive 24, positive 5,
positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send: diag volt iob
Receive: diag volt iob 24.10 4.90 3.20 -3.20
diag volt mb
This command reports the diagnostic voltage measurements on the
motherboard. The sequence of voltages is: Positive 24, positive 15, positive
5, positive 3.3, and negative 3.3. Each voltage value is separated by a space.