Seitenansicht 278
System Description
Electronics
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 8-7
External connectors include:
External Accessory
RS-232/485 Communications (two connectors)
Ethernet Communications
I/O connector with Power Fail Relay, 16 Digital Inputs, and 6
Analog Voltage Outputs.
Internal connectors include:
Function Key Panel and Display
Measurement Interface Board
I/O Expansion Board
Digital Output Board
AC Distribution
The nephelometer interface board serves as the circuitry for the
nephelometer measurement, an internal valve control, and on board
processing of zero and particle concentration by photometry.
Connectors located on the nephelometer interface board include:
RS485
24 VDC Supply
LED Source Current
Reference Detector
Scattering Detector
T/RH Sensor
The measurement interface board serves as a central connection area for all
measurement electronics in the instrument. It contains power supplies and
interface circuitry for sensors and control devices in the measurement
system. It sends status data to the motherboard and receives control signals
from the motherboard.
Nephelometer
Interface Board
Nephelometer interface
board connectors
Measurement Interface
Board